- The column switching-LC/MS permits direct injection of the
sample, thus avoiding contamination in pretreatment. It also
facilitates quantitation because the NP isomers give a single
peak on the chromatogram.
- The nonyl group in NP isomers is straight or branched.
Commercially available standard NP products accordingly consist
of either the straight form only or a mixture of various
branched forms. The mixed type is preferable for the present
purpose, since environmental and biological samples usually give
peak patterns corresponding to them. It should be noted,
however, the peak patterns of the commercial products may vary
among manufacturers, or even among lots from a single
manufacturer.
- Commonly used substances for correction in the mass
spectrometry of NP are straight-chain 4-n-NP (with deuterium and
13C) and 4-(1-methyl)octylphenol-d5. Since differences in
physical and chemical properties of the straight- and
branched-chain molecules may present problems in correction, the
two substances were compared and the latter was chosen here.
Since the retention time for 4-(1-methyl)octylphenol-d5 is
different from that for NP, the correction is still less than
perfect. Future development of mixed standards with deuterium
and 13C is desired.
- E.g. Tosoh TSK-Precolumn BSA/ODS-S (4.6 x 10 mm, 5 ƒÊm).
Other products with equivalent recovery ratio and purification
performance may be used.
- The ƒÀ-glucuronidase concentration is not restricted to this
value. However, the deconjugation condition should always be
checked beforehand.
- Direct injection with column switching is the preferred
method for small numbers of samples. Since the recovery ratio
and purification performance of the extraction column are
effective for about 20 samples only, Procedure B is preferable
for analysis of a large number of samples.
- E.g. Geiman Japan's P/N Ekicrodisc 13 CR 0.45 ƒÊm PTFE.
- E.g. Kanto Chemical's Mightysil RP 18 GP.
- The concentration of ammonium acetate is so chosen that the
maximum response is obtained in a range of 0 to 2 mM. The
solutions were prepared here so that both water and acetonitrile
contain 0.02% ammonium acetate.
- Up to 10% methanol may be contained for column cleaning. The
extraction ratio should then be checked.
- The definitions of the detection and determination limits
follow those given in the Manual for Water Quality Survey
Related to Dioxins, Environment Agency, July 1998.
|